Product Model
|
TRED30-16R (Replaceable Battery).
|
External Sensor Measurement Range
|
-40°C to +99°C (-40°F to +210°F).
|
Operating Temperature Range
|
-30°C to +60°C (-22°F to +140°F).
|
Storage Temperature Range
|
-10°C to +40°C (14°F to +104°F).
|
Rated Temperature Reading Accuracy
|
Better than ±0.5°C (±0.9°F) for -20°C to +40°C (-4°F to +104°F), typically ±0.3°C (0.6°F).
Better than ±0.8°C (±1.4°F) for -40°C to -20°C (-40°F to -4°F), typically ±0.5°C (0.9°F).
Better than ±0.8°C (±1.4°F) for +40°C to +70°C (+104°F to +158°F), typically ±0.7°C (1.3°F).
Better than ±1.2°C (±2.2°F) for +70°C to +99°C (+158°F to +210°F), typically ±1°C (1.8°F).
Actual performance is typically much better than the rated values. Please see the Rated Absolute Accuracy chart below.
Accuracy figures can be improved by recalibration.
|
Rated Temperature Reading Resolution
|
0.1°C (0.2°F) for measurements -40°C to +50°C (-40°F to +122°F).
0.2°C (0.4°F) for measurements +50°C to +70°C (+122°F to +158°F).
0.3°C (0.5°F) for measurements +70°C to +80°C (+158°F to +176°F).
0.4°C-0.6°C (0.7°F to 1°F) for measurements +80°C to +99°C (+176°F to +210°F).
Please see the Rated Native Resolution chart below.
LogTag Analyzer® currently displays to one decimal place of °C or °F. The native resolution is what is stored in the LogTag®.
|
Sensor Reaction Time
|
Typically less than 2 minutes (T90) in moving air (1m/s) for ST100T, ST100H and ST100S types.
|
Recording Capacity
|
15,905 temperature readings. 66 days @ 6min logging.
Day summary statistics memory (for display on LCD): up for 30 days of Max/Min/duration and alarm trigger statistics.
|
Sampling Interval
|
Configurable from 30 seconds to hourly.
|
Logging Start Options
|
Push button start or specific date & time. Optional start delay of up to 18 hours.
|
Recording Indication
|
State indicator “RECORDING".
|
Download Time
|
Typically less than 10 seconds for full memory (15,905 readings), depending on computer or readout device used.
|
Environmental
|
IP61 (when vertically mounted or hung).
|
Power Source
|
3V CR2032 (LiMnO2) coin cell battery, user replaceable.
|
Battery Life
|
Typically 1 year of operation.
NOTE: Typical battery life assumes 6 minute logging, day statistics are reviewed on the display no more than once daily
for no longer than 30 seconds each time.
Download data monthly.
NOTE: Continual activation of the audible alarm will reduce the working life of the battery. When an alarm is triggered,
the alarm should be cleared and reset as soon as possible.
|
Real Time Clock
|
Built-in real time clock.
Rated accuracy ±25ppm @ 25°C (equivalent to 2.5 seconds/day).
Rated temperature coefficient is -0.034 ±0.006ppm/°C (I.e typically +/- 0.00294 seconds/day/°C).
|
Standard Remote Sensor Cable Length
|
Standard: 1.5m (4’11").
Extended: 3m (9’10") (Recommended maximum).
|
Remote Sensor Cable Type
|
PTFE (FDA food contact rated), coaxial.
|
Size
|
93mm(H) x 54.5mm(W) x 8.6mm(T).
|
Weight
|
41g.
|
Case Material
|
Polycarbonate.
|
Compliance and Certifications
|
|