TRED30-16R Display Temp. Logger

Product description

The TRED30-16R supersedes the existing TRED30-7 model. The LogTag® TRED30-16R Temperature Logger measures and stores up to 15,900 temperature readings over a -40°C to +99°C (-40°F to +210°F) measurement range from an External Probe.

Download

Product classification

Product Model
TRED30-16R (Replaceable Battery).
External Sensor Measurement Range
-40°C to +99°C (-40°F to +210°F).
Operating Temperature Range
-30°C to +60°C (-22°F to +140°F).
Storage Temperature Range
-10°C to +40°C (14°F to +104°F).
Rated Temperature Reading Accuracy
Better than ±0.5°C (±0.9°F) for -20°C to +40°C (-4°F to +104°F), typically ±0.3°C (0.6°F).
Better than ±0.8°C (±1.4°F) for -40°C to -20°C (-40°F to -4°F), typically ±0.5°C (0.9°F).
Better than ±0.8°C (±1.4°F) for +40°C to +70°C (+104°F to +158°F), typically ±0.7°C (1.3°F).
Better than ±1.2°C (±2.2°F) for +70°C to +99°C (+158°F to +210°F), typically ±1°C (1.8°F).
Actual performance is typically much better than the rated values. Please see the Rated Absolute Accuracy chart below.
Accuracy figures can be improved by recalibration.
Rated Temperature Reading Resolution
0.1°C (0.2°F) for measurements -40°C to +50°C (-40°F to +122°F).
0.2°C (0.4°F) for measurements +50°C to +70°C (+122°F to +158°F).
0.3°C (0.5°F) for measurements +70°C to +80°C (+158°F to +176°F).
0.4°C-0.6°C (0.7°F to 1°F) for measurements +80°C to +99°C (+176°F to +210°F).
Please see the Rated Native Resolution chart below.
LogTag Analyzer® currently displays to one decimal place of °C or °F. The native resolution is what is stored in the LogTag®.
Sensor Reaction Time
Typically less than 2 minutes (T90) in moving air (1m/s) for ST100T, ST100H and ST100S types.
Recording Capacity
15,905 temperature readings. 66 days @ 6min logging.
Day summary statistics memory (for display on LCD): up for 30 days of Max/Min/duration and alarm trigger statistics.
Sampling Interval
Configurable from 30 seconds to hourly.
Logging Start Options
Push button start or specific date & time. Optional start delay of up to 18 hours.
Recording Indication
State indicator “RECORDING".
Download Time
Typically less than 10 seconds for full memory (15,905 readings), depending on computer or readout device used.
Environmental
IP61 (when vertically mounted or hung).
Power Source
3V CR2032 (LiMnO2) coin cell battery, user replaceable.
Battery Life
Typically 1 year of operation.

NOTE: Typical battery life assumes 6 minute logging, day statistics are reviewed on the display no more than once daily 

for no longer than 30 seconds each time.

Download data monthly.

NOTE: Continual activation of the audible alarm will reduce the working life of the battery. When an alarm is triggered, 

the alarm should be cleared and reset as soon as possible.

Real Time Clock
Built-in real time clock.
Rated accuracy ±25ppm @ 25°C (equivalent to 2.5 seconds/day).
Rated temperature coefficient is -0.034 ±0.006ppm/°C (I.e typically +/- 0.00294 seconds/day/°C).
Standard Remote Sensor Cable Length
Standard: 1.5m (4’11").
Extended: 3m (9’10") (Recommended maximum).
Remote Sensor Cable Type
PTFE (FDA food contact rated), coaxial.
Size
93mm(H) x 54.5mm(W) x 8.6mm(T).
Weight
41g.
Case Material
Polycarbonate.
Compliance and Certifications

FDA CFR21 Part 11 FCC EMC Directive CE EMC Directives Conforms to RoHS




ACCESSORIES




Interface Cradle
External Probe - ST100 Serial Wall Mount