Product Model
|
UTRID-16 (Replaceable Battery).
|
Sensor Measurement Range
|
-30°C to +60°C (-22°F to +140°F).
|
Operating Temperature Range
|
-30°C to +60°C (-22°F to +140°F).*
|
Storage Temperature Range
|
0°C to +40°C (32°F to +104°F).
|
Rated Temperature Reading Accuracy
|
Better than ±0.5°C (±0.9°F) for -20°C to +40°C (-4°F to +104°F).
Better than ±0.7°C (±1.3°F) for -25°C to -20°C (-13°F to -4°F) & +40°C to +60°C (+104°F to +140°F).
Actual performance is typically much better than the rated values. Please see the Rated Absolute Accuracy chart below.
Accuracy figures can be improved by recalibration.
|
Rated Temperature Reading Resolution
|
Less than 0.1°C/F.
Please see the Rated Native Resolution chart below.
LogTag Analyzer® currently displays to one decimal place of °C or °F. The native resolution is what is stored in the LogTag.®
|
Sensor Reaction Time
|
Typically less than 7 minutes (T90) in moving air (1m/s) in EN12830:1999.
|
Recording Capacity
|
16,129 temperature readings. 56 days @ 5min logging, 168 days @ 15min logging.
Statistics memory (for display on LCD):Max/Min values and alarm duration values.
|
Sampling Interval
|
Configurable from 30 seconds to 18 hours.
|
Logging Start Options
|
Push button start (with configurable start delay from 1 minute to 72 hours) or specific date & time.
|
Recording Indication
|
“REC” symbol on LCD.
|
Download Time
|
Typically with full memory (16,129 readings) in less than 30 seconds from time of insertion to availability of PDF report.
Typically less than 10 seconds from time of insertion to availability of LTD file in LogTag® Analyzer (if configured).
|
Environmental
|
IEC 60529: IP64 with USB cap fitted.
|
Power Source
|
3V CR2032 Battery (User replaceable).
|
Battery Life
|
Typically 1 year of operation with normal use (6 minute logging, statistics reviewed on the display no more than once
daily for no longer than 30 seconds each time, download data monthly) provided the logger is kept within the storage temperature
range when not in use.
The battery low symbol will appear on the LCD if the UTRID-16 has a low battery.
|
Real Time Clock
|
Built-in real time clock.
Rated accuracy ±25ppm @ 25°C (equivalent to 2.5 seconds/day).
Rated temperature coefficient is -0.034 ±0.006ppm/°C (I.e typically +/- 0.00294 seconds/day/°C).
|
Connection Interface
|
USB 2.0, A-type plug.
|
Software
|
PDF Reader, LogTag® Analyzer 2.8 or higher.
|
Size
|
93mm(H) x 54.5mm(W) x 8.6mm(T) including protective USB cap.
|
Weight
|
39g.
|
Case Material
|
Polycarbonate.
|
Footnote
|
*(-30°C to +60 °C applies to Loggers with serial numbers 6x45x15811 or greater. Loggers with serial numbers prior to this have an
operating temperature range of -25°C to +60°C).
|