USRIC-4 USB Temperature Logger

Product description

Designed as a Single-use, cost effective & reliable temperature recorder, the LogTag® USRIC-4 measures and stores real time temperature readings over a measurement range of -30°C to +60°C (-22°F to +140°F). The USRIC-4 combines all the features of an advanced temperature recorder with a 3937 log memory and a 6-month operating life in a robust and inexpensive package.

Download

Product classification

Product Model
USRIC-4
Sensor Measurement Range
-30°C to +60°C (-22°F to +140°F)
Operating Temperature Range
-30°C to +60°C (-22°F to +140°F).*
Storage Temperature Range
0°C to +40°C (-32°F to +104°F).
Rated Temperature Reading Accuracy
Better than ±0.5°C for -5°C to +30°C.
Better than ±0.8°C or better for measurements in other areas of rated range.
Actual performance is typically much better than the rated values. Please see the Rated Absolute Accuracy chart below.
Accuracy figures can be improved by recalibration.
Rated Temperature Reading Resolution
Less than 0.1°C/F.
Please see the Rated Native Resolution chart below.
LogTag Analyzer® currently displays to one decimal place of °C or °F. The native resolution is what is stored in the LogTag®.
Sensor Reaction Time
Typically less than 5 minutes (T90) in moving air (1m/s).
Recording Capacity
3937 temperature readings. 14 days @ 5min logging, 41 days @ 15min logging.
Sampling Interval
Configurable from 1 minute to 18 hours.
Logging Start Options
Push button start or specific date & time.
Recording Indication
Flashing ‘OK’ indicator / flashing ‘ALERT’ indicator.
Download Time
Typically less than 4 seconds for full memory (3937 readings), depending on computer or readout device used.
Environmental
IP64 with USB cap fitted.
Power Source
3V Li-Mg Battery.
Battery Life
Shelf life of up to two years before configuration. Six months operating life from the time of configuration.
Real Time Clock
Built-in real time clock.
Rated accuracy ±25ppm @ 25°C (equivalent to 2.5 seconds/day).
Rated temperature coefficient is -0.034 ±0.006ppm/°C (I.e typically +/- 0.00294 seconds/day/°C).
Connection Interface
USB 2.0, A-type plug
Software
LogTag® Analyzer 2.5 or higher
Size
93mm(H) x 54.5mm(W) x 8.6mm(T).
Weight
30g.
Case Material
Polycarbonate.
Footnote
*(-30°C to +60 °C applies to Loggers with serial numbers 5x34x33580 or greater. Loggers with serial numbers prior to this have an operating temperature range of -25°C to +60°C).
Compliance and Certifications

FDA CFR21 Part 11 FCC EMC Directive CE EMC Directives Conforms to RoHS



ACCESSORIES



Protective Enclosure Wall Mount