Product Model
|
Standard Model: TIC20
Extention Model: WHO-Type 1, WHO-Type 2, WHO-Type P and WHO-Type R
|
Operating Temperature Range
|
-30°C to +60°C
|
Storage Temperature Range
|
-30°C to +65°C
|
Ambient humidity range during transport and use
|
0 to 95%RH
|
Resolution
|
0.1°c or better in range of -30°C to +60°C
|
Accuracy
|
• Better than ±0.5°C for -10°C to +40°C
• Better than ±1.0°C for -30°C to -10°C
• Better than ±1.0°C for +40°C to +60°C
|
Capacity
|
• Minimum & Maximum temperature for each of the 20 days
• 1st activation of each Alarm for each day including Trigger time and duration of each excursion
|
Memory type
|
Non volatile
|
Sampling Interval
|
Factory set to 6 minute interval (i.e 0.1hour)
|
Logging modes
|
20 day statistics, temperature/time excursions and Alarm trigger time and duration for each day
|
Logging Start delay
|
Factory set to 60 minute interval
|
Sensor
|
Precision thermistor
|
Sensor Reaction time
|
T90 less than 7 minutes by method detailed in EN12830:1999
|
Vibration
|
Withstands vibration specification as detailed in EN12830:1999
|
Shock
|
• Withstands shock specification as detailed in EN12830:1999
• Withstands 5 drops from 1m to smooth concrete floor without loss of function or calibration
|
Environmental
|
IEC 60529: IP64
|
Power source
|
3V Lithium battery – non replaceable
|
Battery life
|
• Minimum storage life of 18 Months before ‘start’
• Monitoring period: 20 days
• Minimum accessibility (display) period of 6 months after ‘stopped’
|
Size
|
73mm(H) x 54.5mm(W) x 8.6mm(T) (Volume < 34cm3)
|
Weight
|
30g
|
Case Material
|
Polycarbonate
|
Calibration
|
Covered by Certificate of Traceability. Calibration traceable to an ISO/IEC 17025 accredited testing laboratory.
|
EMC compliance
|
• EC EMC directives (EN 61000-6-1:2005 & EN 61000-6-3:2006)
• Includes electrostatic discharge as prescribed in EN 61000-4-2
• Complies with FCC Part 15 Subparts A and B
|
Alarms function & display
|
• Alarm functions as prescribed in the WHO PQS performance specification E06/TR07.2
• Alarms are displayed on a custom LCD
• Display activates when any button is pressed
• Display is static allowing photocopying
|