Product Model
|
TRIL-8 (Multi-Trip)
|
Sensor Measurement Range
|
-80°C to +40°C (-112°F to +104°F).
|
Operating Temperature Range
|
-80°C to +40°C (-112°F to +104°F).
|
Storage Temperature Range
|
-20°C to +40°C (-4°F to +104°F).
|
Rated Temperature Reading Accuracy
|
Better than ±1°C for -30°C to +20°C.
Better than ±1.2°C for -45°C to -30°C and +20°C to +40°C.
Better than ±1.7°C for -80°C to -45°C.
Actual performance is typically much better than the rated values. Please see the Rated Absolute Accuracy chart below.
Accuracy figures can be improved by recalibration.
|
Rated Temperature Reading Resolution
|
Less than 0.1°C for -80°C to 0°C.
Less than 0.2°C for 0°C to +20°C.
Less than 0.5°C for +20°C to +40°C.
Please see the Rated Native Resolution chart below.
LogTag Analyzer® currently displays to one decimal place of °C or °F. The native resolution is what is stored in the LogTag®.
|
Sensor Reaction Time
|
Typically less than 5 minutes (T90) in moving air (1m/s).
|
Recording Capacity
|
8000 temperature readings. 53 days @ 10min logging, 80 days @ 15min logging.
|
Sampling Interval
|
Configurable from 1 minute to 18 hours.
|
Logging Start Options
|
Push button start or specific date & time.
|
Recording Indication
|
Flashing ‘OK’ indicator / flashing ‘ALERT’ indicator.
|
Download Time
|
Typically less than 5 seconds for full memory (8000 readings), depending on computer or readout device used.
|
Environmental
|
IP65 (roughly equivalent to NEMA 4).
|
Power Source
|
3.6V low temperature chemistry Lithium battery.
|
Battery Life
|
Storage life of up to 12 months before ‘start’. Rated for a typical trip of 2 weeks duration at dry ice temperatures.
|
Real Time Clock
|
Built-in real time clock.
Rated accuracy ±25ppm @ 25°C (equivalent to 2.5 seconds/day).
Rated temperature coefficient is -0.034 ±0.006ppm/°C (I.e typically +/- 0.00294 seconds/day/°C).
|
Size
|
86mm(H) x 54.5mm(W) x 8.6mm(T).
|
Weight
|
35g.
|
Case Material
|
Polycarbonate.
|
Compliance and Certifications
|
|